Electromagnetic Compatibility

The PL15T02 PowerLab 15T, PL2602 PowerLab 2/26, PL2604 PowerLab 4/26 and
PL26T04 PowerLab 26T (the devices) have been tested to comply with IEC 60601-1-
2:2014. The relevant basic EMC standards are IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-
6, IEC 61000-4-8, IEC 61000-4-11 and CISPR 11.
Emissions
• The devices are suitable for use in all establishments other than domestic
Immunity
• Mains power quality should be that of a typical commercial or hospital
• Power frequency magnetic fields should be at levels characteristic of a typical
• Floors should be wood, concrete or ceramic tile. If floors are covered with
Separation Distances
• The devices are intended for use in an electromagnetic environment in which
• Portable and mobile RF communications equipment should be used no closer
• Field strengths from fixed RF transmitters, as determined by an electromagnetic
Rated maximum output
power of transmitter, P
45
and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes. There may be potential
diff iculties in ensuring electromagnetic compatibility in other environments, due
to conducted as well as radiated disturbances.
environment. If the user of the device requires continued operation during
power mains interruptions, it is recommended that the device be powered from
an uninterruptible power supply or a battery.
location in a typical commercial or hospital environment.
synthetic material, the relative humidity should be at least 30%.
radiated RF disturbances are controlled.
to any part of the device, including cables, than the recommended separation
distance in the table below.
site survey, should be less than the compliance level in each frequency range.
0.01 W
0.1 W
1 W
10 W
100 W
Separation distance
150 kHz to 800 MHz
d = 1.17√P
0.1 m
0.4 m
1.2 m
3.7 m
11.7 m
PowerLab Teaching Series - Owner's Guide
800 MHz to 2.7 GHz
d = 2.33√P
0.2 m
0.7 m
2.3 m
7.4 m
23.4 m