Output Amplifier
Output configuration:
Output resolution:
Maximum output current:
Accuracy error:
Linearity error:
Output ranges:
Output slew rate:
Settling time (G=1, 10 V step): 5 µs
Output impedance:
Isolated Stimulator Output (PowerLab 15T & 26T)
Output configuration:
Isolation rating:
Pulse duration:
Compliance voltage:
Output current:
Safety indicators:
Safety switch:
External Trigger (not on PowerLab 15T)
Trigger mode:
Trigger threshold:
Hysteresis:
41
2 Outputs- Complementary
16-bit
20 mA
Less than 0.2 5
±0.5 LSB (INL) (typical)
±0.5 LSB (DNL) (typical)
± 10 V
± 5 V
± 2 V
± 1 V
± 500 mV
± 200 mV
2.3 V/µs
0.001 Ω
Constant-current stimulator with hardware limited
repitition rate
4000 V
to ground as per IEC60601-1
rms
2000 V
to Bio Amp inputs
rms
50 - 200 µs (soft ware selectable)
100 - 110 V typical
Soft ware selectable, but hardware limited to a maximum
of 1000 Hz and 200 µs for safety
Green and yellow indicators. Green flash indicates
delivery of a valid stimulus. Concurrent green and yellow
flash indicates an out-of-compliance condition.
Provides physical disconnection of the stimulator from
the subject.
TTL level or contact closure, soft ware selectable
+1.3 V (rising edge), +1.1 V (falling edge)
0.3 V