Gossen MetraWatt SECUTEST BASE Operating Instructions Manual

Gossen MetraWatt SECUTEST BASE Operating Instructions Manual

Tester for measuring the electrical safety of devices per vde 0701-0702, iec 62353 and iec 60974-4
Hide thumbs Also See for SECUTEST BASE:
Table of Contents

Advertisement

Quick Links

Condensed Operating Instructions
SECUTEST
and SECULIFE ST
Tester for Measuring the Electrical Safety of Devices
per VDE 0701-0702, IEC 62353 and IEC 60974-4
B
(
) /
10
ASE
BASE
Read carefully before use.
Keep on file for future reference!
Please read the full operating instructions as well,
which are available as a PDF file at
www.gossenmetrawatt.com.
The condensed operating instructions do not
replace the full operating instructions!
P
RO
(
)
25
Important
3-349-835-03
8/8.18

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SECUTEST BASE and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Gossen MetraWatt SECUTEST BASE

  • Page 1 Condensed Operating Instructions SECUTEST and SECULIFE ST BASE Tester for Measuring the Electrical Safety of Devices 3-349-835-03 per VDE 0701-0702, IEC 62353 and IEC 60974-4 8/8.18 Important Read carefully before use. Keep on file for future reference! Please read the full operating instructions as well, which are available as a PDF file at www.gossenmetrawatt.com.
  • Page 2 SECUTEST (10)/P & SECULIFE ST (25) Operating and Connection Overview BASE Operating and Connection Overview Symbol display for devices Special Symbol Displays: connected to the USB master – Measurement at IT line active Kink protection grommet: interface see below black: max. 16 A –...
  • Page 3: Table Of Contents

    2.2 Detection of Mains Connection Errors ....6 Installed DUTs of Safety Class I ....... 27 Overview of Features 8.17 Measurements with Test Adapter ....28 Included with SECUTEST BASE(10)/PRO and Test Sequences in Accordance with SECULIFE ST BASE Testers ......7 Standards ..........29 User Interface Symbols –...
  • Page 4: Safety Precautions

    Safety Precautions • Make certain that the measurement cables are in flawless condition, e.g. no SECUTEST BASE(10), SECUTEST PRO and damage to insulation, no cracks in SECULIFE ST BASE(25) test instruments are cables or plugs etc. manufactured and tested in accordance with the following safety regulations: •...
  • Page 5 SECUTEST (10)/P & SECULIFE ST (25) Safety Precautions BASE Opening the instrument / repairs The test instrument may not be used: The instrument may only be opened by • If external damage is apparent, for example authorized, trained personnel in order to if parts which conduct dangerous touch ensure flawless operation and to assure that voltage are freely accessible, if the display...
  • Page 6: Initial Start-Up

    SECUTEST (10)/P & SECULIFE ST (25) Initial Start-Up BASE Initial Start-Up Detection of Mains Connection Errors The device automatically recognizes mains Mains Connection connection errors if the conditions in the fol- lowing table have been fulfilled. The user is Nominal mains values: 100 to 240 V, 50 Hz to 400 Hz informed of the type of error, and all measuring ➭...
  • Page 7: Overview Of Features

    SECUTEST (10)/P & SECULIFE ST (25) Scope of Features BASE Overview of Features Included with SECUTEST BASE(10)/PRO and SECULIFE ST BASE Testers Switch Measuring functions, Switch Measuring functions, setting, test current/voltage setting, test current/voltage descrip- descrip- tion as of tion as of...
  • Page 8: User Interface Symbols - Parameter And Softkey Symbols

    SECUTEST (10)/P & SECULIFE ST (25) Symbols BASE User Interface Symbols – Parameter and Softkey Symbols Sym- Setup Parameters and their significance bols Page Complete overviews of all symbols are included in the full operating instructions. All measurements: Ref.voltage L-PE: voltage to which the measured values for leakage current have been standardized;...
  • Page 9: Internal Database

    SECUTEST (10)/P & SECULIFE ST (25) Database – Keyboard BASE Internal Database Data Input Creation of Test Structures Keyboard or Softkey Entries A complete test structure with data regard- After selecting ID or any other object param- ing customers and devices under test be eter, a keyboard is displayed which allows created in the test instrument.
  • Page 10: Notes On Saving Single Measurements And Test Sequences

    SECUTEST (10)/P & SECULIFE ST (25) Saving BASE Notes on Saving Single Measurements and Test Sequences At the end of each test, test results can be Variant 3 – entry of a new ID at the end of the test saved under an ID number which is unequiv- You haven’t yet set up a test structure in the ocally assigned to the respective test object.
  • Page 11: Single Measurements

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE Single Measurements Procedure for measuring with save function and subsequent entry of the (ME) device Any measuring duration is possible. The As an alternative to the procedure described respective measurement is ended by press- above, you can start with step 4 and, after ing START/STOP.
  • Page 12: Rpe - Protective Conductor Resistance For Protection Class I Devices

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE RPE – Protective Conductor Resistance for Protection Class I Devices SECUTEST B 10 / SECUTEST PRO / SECULIFE ST BASE SECULIFE ST BASE25 Select measuring function The selected offset value is permanently stored to memory and adapted for measurements in selector switch position AUTO.
  • Page 13: Riso - Insulation Resistance Measurement For Protection Class I Devices

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE RISO – Insulation Resistance Measurement for Protection Class I Devices Select measuring function Connect DUT RISO Select parameters – LN(TS) PE(TS) Special case: inputs for application parts U+/U– = increase/decrease UINS(set) –...
  • Page 14: Protection Class Ii Devices

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE RISO – Insulation Resistance Measurement for Protection Class II Devices Select measuring function Connect DUT RISO – LN(TS) Select parameters ➭ Connect the DUT to the test socket. ➭ Contact all conductive, exposed parts with test probe P1.
  • Page 15: Ipe - Protective Conductor Current

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE IPE – Protective Conductor Current Select Measuring Function Connect DUT Select parameters Direct / Diff / Alt ➭ Connect the DUT (to test socket). Start test Acknowledge line voltage warning Set parameters Direct &...
  • Page 16: Ib - Touch Current

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE IB – Touch Current Select measuring function Connect DUT Select parameters Direct P1/Diff P1/Alt P1 ➭ Connect the DUT to the test socket. ➭ Contact additional, accessible, conduc- tive parts which are not connected to the protective conductor with test probe P1.
  • Page 17: Ig - Device Leakage Current

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE IG – Device Leakage Current Select measuring function Connect DUT Select parameters Direct / Alt ➭ Connect the DUT to the test socket. ➭ Contact accessible, conductive parts which are not connected to the protec- tive conductor with test probe P1.
  • Page 18: Ia - Leakage Current From The Application Part

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE IA – Leakage Current from the Application Part Select measuring function Connect DUT Direct P1/ PC 1 device under test Select parameters ➭ Connect the DUT (to test socket). ➭ Contact the short-circuited inputs for the application parts with test probe P1.
  • Page 19: Ip - Patient Leakage Current

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE IP – Patient Leakage Current Select measuring function Connect DUT Select parameters ➭ Connect the DUT to the test socket. ➭ Contact the short-circuited inputs for the application parts with test probe P1. Start test Acknowledge line voltage warning Set parameters...
  • Page 20: U - Probe Voltage

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE U – Probe Voltage ➭ Connect the DUT to the test socket. ➭ Contact the ungrounded output for pro- Select measuring function tective extra-low voltage with test probe ➭ Select line voltage polarity. Special case: permanently installed DUT Probe Voltage Select parameters...
  • Page 21: U - Measuring Voltage

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.10 U – Measuring Voltage SECUTEST PRO and SECULIFE ST BASE only: Special case: permanently installed DUT Select measuring function Meas. Volt. V – COM Select parameters ➭ For testing mains power packs and charging units: Connect the DUT mains terminal with the test socket.
  • Page 22: Ta - Prcd Time To Trip (Portable Residual Current Device)

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.11 tA – PRCD Time to Trip (portable residual current device) Select Measuring Function Acknowledge line voltage warning Execute test ➭ Activate the PRCD. ➭ Contact neutral conductor L at the PRCD with test probe P1 (if required, as- certain by trial and error).
  • Page 23: P - Functions Test

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.12 P – Functions Test Select measuring function Connect DUT Select parameters ➭ Connect the DUT to the test socket. Starting test Acknowledge line voltage warning Set parameters Meas. Parameter Meaning Polarity Save measured values to clipboard Phase L –...
  • Page 24: El1 - Testing Extension Cords For Continuity

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.13 EL1 – Testing Extension Cords for Continuity, Short-circuit and Polarity Select measuring function Connect DUT EL1-Adapter Test Socket – P1 Select parameters VL2E-adapter Connection of EL1 Adapter ➭ Connect the EL1 adapter to the P1 Set parameters probe sockets at the test instrument.
  • Page 25: Extra - Special Functions

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.14 EXTRA – Special functions SECUTEST BASE(10) SECUTEST PRO (Feature I01) & SECULIFE ST BASE(25) Select measuring function Select measuring function EXTRA EXTRA In this case, additional measurement functions are assigned to rotary switch position EXTRA.
  • Page 26 SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE Temp. – Temperature measurement IZ – Current clamp measurement Connect DUT Connect DUT Temperature measurement works both with a Pt100 and a Pt1000 temperature sensor Set parameters and automatically recognizes the utilized Meas.
  • Page 27: 2-Pole Measurements With P1 And P2 Test Probes

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.15 2-Pole Measurements with P1 and P2 8.16 Measurement with Current Clamp Sensor at Test Probes Permanently Installed DUTs of Safety Class I SECUTEST PRO and SECULIFE ST BASE(25) only SECUTEST PRO and SECULIFE ST BASE(25) only In case your DUT is not equipped with a SECUTEST PRO Clamp...
  • Page 28: Measurements With Test Adapter

    SECUTEST (10)/P & SECULIFE ST (25) Single Measurements BASE 8.17 Measurements with Test Adapter Connection Example with EL1 Test with Adapter EL1 VL2E AT3- AT16DI CEE- IIIE AT32DI Adapter EL1 adapter DUT terminals ✔ ✔ Inlet plug 1P+N+PE 16 A —...
  • Page 29: Test Sequences In Accordance With Standards

    Test Sequences in Accordance with Standards Test Sequences in Accordance 9 At the end of the test sequence, you can generate a list of the with Standards results of the individual test steps. If the same sequence of single tests will be 10 If you want to view details such as run frequently (one after the other with sub- the settings for the individual test...
  • Page 30: Sample Test Sequence

    SECUTEST (10)/P & SECULIFE ST (25) Test Sequences BASE Sample Test Sequence Sequence parameters Select test sequence Individual test steps can be configured with the sequence parameters, see detailed operating instructions. Set classification parameters Switch setting: A1 ... A8 Meas. Parameter Meaning As-delivered Condition: Protection class *...
  • Page 31 SECUTEST (10)/P & SECULIFE ST (25) Test Sequences BASE Connect DUT Manual evaluation of visual inspection ➭ Connect the DUT to the test instrument in accordance with the selected test se- quence. – Test socket – Permanent connection – Adapter Switch position: A1 ...
  • Page 32 SECUTEST (10)/P & SECULIFE ST (25) Test Sequences BASE Optional test step Test step – automatic evaluation End of sequence – display results (display of the memory screen depends on the The measured value is ascertained automat- parameter pre-selection in the SETUP switch po- ically within a specified period of time.
  • Page 33: Measurement Parameters For Individual Measurements And Test Sequences

    Measurement Parameters SECUTEST (10)/P & SECULIFE ST (25) BASE Optional test step Measurement Parameters for Individual Measurements and Confirm results Test Sequences Measurement parameters which apply for both individual measurements and test sequences, have to be entered in selector switch position SETUP. Setup 1/3 >...
  • Page 34: Repair And Replacement Parts Service Calibration Center And Rental Instrument Service

    SECUTEST (10)/P & SECULIFE ST (25) Addresses BASE Repair and Replacement Parts Service Product Support Calibration Center and Rental Instru- If required please contact: ment Service If required please contact: GMC-I Messtechnik GmbH Product Support Hotline Phone, +49 911 8602-0 GMC-I Service GmbH Fax: +49 911 8602-709...
  • Page 35: Report Software

    SECUTEST (10)/P & SECULIFE ST (25) BASE Report Software IZYTRONIQ is a test software that has been newly developed from scratch. It enables the user to visualize and manage the entire test- ing procedure for all our test instruments and to document it in an audit-proof manner. For the first time, it is thus possible to com- bine the test and measurement data from a great variety of test instruments and multi-...
  • Page 36 SECUTEST (10)/P & SECULIFE ST (25) BASE Edited in Germany • Subject to change without notice • PDF version available on the Internet Phone: +49 911 8602-111 GMC-I Messtechnik GmbH Fax: +49 911 8602-777 Südwestpark 15 e-mail: info@gossenmetrawatt.com 90449 Nürnberg, Germany www.gossenmetrawatt.com...

Table of Contents