Table 2: Attendant Control (1742) Diagnostics - Curtis enAble 40 Manual

Powerchair control system
Table of Contents

Advertisement

Curtis enAble
40 Manual,
®
ATTENDANT CONTROL DIAGNOSTICS
The LEDs on the attendant control are used to communicate information
about faults, warnings, and battery charging.
The Mode LED, the BDI LED, and the LEDs in the chair icon signal
warnings by flashing. For more information, turn on the user handcontrol and
refer to Table 1.
Table 2: ATTENDANT CONTROL FAULT & WARNING INDICATORS
SIGNAL
MODE LED
(RED)
(Various faults and warnings.)
CHAIR ICON LEDs
(GREEN)
Flashing
Seat Recline Fault.
Seatback LED
Flashing
Seat Lift Fault.
Seat LED
Flashing
Left or Right Leg Fault.
Leg LED
All the chair
Actuator Over/Under Temperature
LEDs flashing
Warning.
BDI LED
Flashing Red
Undervoltage.
Flashing Green
Overvoltage.
Flashing Red,
Charging battery; inhibit.
Amber, Green
in Sequence
Rev. E
8 — DIAGNOSTICS & TROUBLESHOOTING
FAULT/WARNING/INFO
COMMENTS
See display on the user
handcontrol to ascertain
which fault is present; then
see Table 1 for further
information.
See Table 1.
See Table 1.
See Table 1.
See Table 1.
See Table 1.
See Table 1.
51

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents