Principal Switch Operations; A-Bpm Operations - A&D TM-2440 Instruction Manual

Recorder for ambulatory blood pressure monitor
Hide thumbs Also See for TM-2440:
Table of Contents

Advertisement

5.3. Principal Switch Operations

5.3.1. A-BPM Operations

To start or suspend A-BPM.
Step 1. Store the preset program (of start times and intervals) for A-BPM.
Step 2. Press and hold the EVENT switch for 3 seconds or more to
switch between the following states.
"ON" ....... A-BPM is started and the
"OFF" ...... A-BPM is suspended and the
To expand A-BPM interval time.
Step 1. Set the sleep mode to "ON" before the measurement.
Step 2. Start A-BPM by pressing and holding the EVENT switch for 3
seconds or more.
The
Step 3. When the EVENT switch is pressed during A-BPM, the interval
time is doubled.
When the EVENT switch is pressed again, the interval time
returns to basic value.
Blood pressure measurements are performed in
accordance with preset A-BPM program.
Manual blood pressure measurement can be
performed by pressing the START switch, though.
mark is shown.
mark is shown.
17
mark turns off.