8.6. Blood Pressure Measurement Operations

8.6.1. A-BPM Operations

When A-BPM is started, blood pressure is measured in
accordance with the preset parameters.
Set the built-in clock and initial pressurization value before
measurement because A-BPM uses them. Refer to "8.2.2. The
Clock and the Monitor Function of Measurement" and "8.3.
A-BPM Preset Programs".
When the recorder is removed, suspend A-BPM by pressing and
holding the EVENT switch for 3 seconds or more.
If the recorder is removed during A-BPM, the inflation of the cuff
is started next start time, the cuff may break.
To resume A-BPM, press and hold the EVENT switch for 3
seconds or more again.
The
mark is displayed while A-BPM is activated.
Manual blood pressure measurement can be performed during
A-BPM waiting mode.
The measurement result of the manual blood pressure
measurement can be stored in the memory.
When A-BPM is stopped, the error code E07 is displayed on
the OLED display and stored in the memory.
To start A-BPM
Step 1. Press and hold the EVENT switch for 3 seconds or more.
Step 2. The
started.
mark is shown on the OLED display. A-BPM is
Note
51