A&D TM-2440 Instruction Manual page 91

Recorder for ambulatory blood pressure monitor
Hide thumbs Also See for TM-2440:
Table of Contents

Advertisement

Immunity test
Radiated RF
IEC61000-4-3
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Note : U
is the AC mains voltage prior to application of the test level.
T
IEC60601-1-2
test level
3 V/m
80 MHz to
2.5 GHz
<5 % U
T
(>95 % dip in U
for 0.5 cycle
<40 % U
T
(>60 % dip in U
for 5 cycle
<70 % U
T
(>30 % dip in U
for 25 cycle
<5 % U
T
(>95 % dip in U
for 5 seconds
Compliance
level
10 V/m
80 MHz to
2.6 GHz
)
T
)
T
N.A.
)
T
)
T
73
Electromagnetic
environment
Recommended
separation distance :
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Does not apply
because the power
supply unit is built in.