To Stop during A-BPM
When the START/STOP switch is pressed during the blood
pressure measurement, the air is exhausted immediately and the
current measurement is stopped. However, A-BPM is continued.
The next blood pressure measurement is performed in
accordance with A-BPM settings.
To set the program for A-BPM.
Step 1. If the indication of the display is hidden, press the
START/STOP or EVENT switch to return to the display of
waiting mode.
Step 2. If the
3 seconds or more to suspend A-BPM.
Step 3. While pressing and holding the START/STOP switch,
press and hold the EVENT switch for 3 seconds or more.
Sleep is displayed on the OLED display.
Step 4. Operation switches are as follows:
Refer to "8.3.1. A-BPM Items and Parameters"
EVENT switch ................. Change the current parameter.
START/STOP switch ...... Decision, next item, end of settings.
To measure blood pressure during A-BPM immediately.
( Manual blood pressure measurement of A-BPM )
Step 1. If the indication of the OLED display is hidden, press the
START/STOP or EVENT switch to return to the display of
A-BPM waiting mode. A-BPM waiting mode is a state that
blood pressure is not measured during the interval time.
Step 2. Press the START/STOP switch during A-BPM waiting mode.
mark is shown, press and hold the EVENT switch for
18