Leica iCON iCR70 User Manual page 59

Table of Contents

Advertisement

Technical Data
Angular accuracy
*
Orthogonal to the line of sight.
Refer to the data sheet of the respective instrument model for infor-
mation about the angular accuracy.
EDM accuracy
The distance measurement accuracy consists of two parts: a fixed value and a
distance-dependent value (ppm-value).
Example: "Single measurements: 1 mm + 1.5 ppm"
The EDM accuracies for prism and reflectorless measurements can differ. Addi-
tionally, the accuracies can differ depending on the used technologies.
Refer to the appropriate data sheet for information about the EDM
accuracy.
ATR accuracy
Automatic target aiming accuracies, like those of the ATR, are in general the
same as the stated angular accuracy. Therefore these accuracies are also dis-
tance-dependent parameters.
External impacts, like heat shimmer, rain (prism surface covered by rain drops),
fog, dust, strong background lights, dirty targets, alignment of the targets etc.
can have a significant influence on the automated target. In addition the selec-
ted EDM mode affects the ATR performance. Under good environmental condi-
tions and with a clean, properly aligned target the accuracy of the automated
target aiming is equivalent to the manual target aiming (presumed valid cali-
bration values).
Type and centring accuracy of the prism
The prism centring accuracy depends mainly on the used prism type, for exam-
ple:
Prism type
Leica GPR1
Leica GPH1P
Leica GRZ122
Leica GRZ4
Refer to the white paper "Leica Surveying Reflectors" for information
about the different centring accuracies.
Additional influencing factors
When determining absolute coordinates, the following parameters can also
affect the resulting accuracy:
Environmental conditions: temperature, air pressure and humidity
Typical instrument errors, such as horizontal collimation error or index
error.
Proper functioning of laser plummet or optical plummet
Correct horizontal levelling
Setup of the target
Quality of additional equipment, such as tribrach or tripod.
Possible deviation
~2.5 mm
Circular prism
Precision circular prism
360° prism
360° prism
*
at 100 m distance
Centring accuracy
1.0 mm
0.3 mm
2.0 mm
5.0 mm
59

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Icon icr80

Table of Contents