Calibration With Any Buffers - JUMO 202710/20 Operating Manual

Handheld meter for ph, redox, and rh
Table of Contents

Advertisement

6 Calibration of the pH combination electrode

6.4.2 Calibration with any buffers

Calibration point 1
rH
mV
mV
pH
H
ATC
ok
CAL
AL Logg
pH
Use the
and
solution and then save by pressing the
Finally, the condition of the combination electrode is briefly displayed and then
the 3 point calibration is completed and the device returns to measurement
mode.
NOTE!
For a sufficiently accurate reading of the slope, the pH values of the selected
buffer solutions should not be too close together (minimum gap of 2 pH from
the neutral buffer).
Press the
key for at least 2 seconds to start the calibration.
Use the
and
Insert the combination electrode into the first (neutral) buffer solution and
gently stir.
NOTE!
If the 2 or 3 point calibration has been selected as the calibration method,
the device expects a neutral buffer solution within the range pH 6.5 to pH 7.5
for the first calibration point.
If the 1 point calibration has been selected as the calibration method, any
buffer solution with a known pH value can be employed.
rH
mV
mV
pH
H
ATC
ok
CAL
AL Logg
pH
Press the
key when the measured value is stable (small triangle
appears on the display above the yellow "ok" label).
The display changes so that the temperature value for
the third buffer solution can be entered.
keys to set the temperature value for the third buffer
keys to set the value for the neutral buffer solution.
The set value for the neutral buffer solution is on the
main display of the device, "Pt.1" and a revolving
symbol appear on the subsidiary display of the
device.
key.
27