Inspection of the endoscopic system

Inspection of the endoscopic image
Inspection of the remote switches
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
Do not stare directly into the distal end of the endoscope
while the examination light is ON. Otherwise, eye injury may
result.
1.
Turn the video system center, light source, and monitor ON and inspect the
WLI and NBI endoscopic image as described in their respective instruction
manuals.
2.
Confirm that light is outputting from the endoscope's distal end.
3.
While observing the palm of your hand, confirm that the WLI and NBI
endoscopic image is free from noise, blur, fog, or other irregularities.
4.
Angulate the endoscope and confirm that the WLI and NBI endoscopic
images do not momentarily disappear or display any other irregularities.
If the object cannot be seen clearly, wipe the objective lens
using a clean, lint-free cloth moistened with 70% ethyl or
70% isopropyl alcohol.
All remote switches should be checked to work normally
even if they are not expected to be used. The endoscopic
image may freeze, or other irregularities may occur during
examination and may cause patient injury, bleeding, and/or
perforation.
Depress every remote switch and confirm that the specified functions work
normally.
Chapter 3 Preparation and Inspection
37