Olympus EVIS EXERA II Operation Manual page 63

Duodenovideoscope
Hide thumbs Also See for EVIS EXERA II:
Table of Contents

Advertisement

Figure 4.6
Guidewire-locking
grove
Figure 4.7
4.
The guidewire is locked at the endoscope's distal end.
5.
Withdraw the EndoTherapy accessory slowly while holding the elevator
control lever stationary so that the elevator and guidewire do not move
forward to the "
while withdrawing the accessory.
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
Guidewire
The stiff part of the guidewire is locked at the
guidewire-locking groove more effectively.
The assist function of the guidewire locking works more
effectively when the papilla is observed on the left side of
endoscopic image.
U" direction. Observe the endoscopic and X-ray images
Chapter 4 Operation