Olympus EVIS EXERA II Operation Manual page 90

Duodenovideoscope
Hide thumbs Also See for EVIS EXERA II:
Table of Contents

Advertisement

Appendix
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Definition:
86
 Guidance and manufacturer's declaration —
Electromagnetic immunity
This model is intended for use in the electromagnetic environment specified
below. The customer or the user of this model should assure that it is used
in such an environment.
Portable and mobile RF communications equipment should be used no
closer to any part of this model, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
IEC 60601 test level
3 Vrms
3 V (V
(150 kHz – 80 MHz)
3 V/m
3 V/m (E
(80 MHz – 2.5 GHz)
Where "P" is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and "d" is the recommended separation distance in meters (m).
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
Compliance
Electromagnetic environment — Guidance
level
Recommended separation distance
)
1
d
=
Recommended separation distance
)
1
d
=
d
=
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Electromagnetic interference may occur in the vicinity of
high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
3,5
------ -
P
V
1
3,5
------ -
P
E
1
80 MHz – 800 MHz
7
----- -
P
E
1
800 MHz – 2.5 GHz