Transportation Of The Endoscope - Olympus EVIS EXERA II Operation Manual

Duodenovideoscope
Hide thumbs Also See for EVIS EXERA II:
Table of Contents

Advertisement

4.4

Transportation of the endoscope

Transporting within the hospital
Transporting outside the hospital
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
When carrying the endoscope by hand, loop the universal cord, hold the
endoscope connector with the control section in one hand and hold the distal
end of the insertion tube securely, but gently without squeezing, in the other
hand (see Figure 4.9).
Figure 4.9
Transport the endoscope in the carrying case.
Always clean, disinfect, or sterilize the endoscope after
removing it from the carrying case. If the endoscope is not
cleaned, disinfected, or sterilized, it could pose an infection
control risk.
The carrying case cannot be cleaned, disinfected, or
sterilized. Clean and disinfect or sterilize the endoscope
before placing it in the carrying case.
To avoid damage to the endoscope caused by changes in air
pressure, do not attach the water-resistant cap when
transporting the endoscope.
Before putting the endoscope in the carrying case, always
make sure that the forceps elevator is not raised. Putting the
endoscope in the carrying case while the forceps elevator is
raised could damage the endoscope.
Chapter 4 Operation
63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents