Olympus EVIS EXERA II Operation Manual page 49

Duodenovideoscope
Hide thumbs Also See for EVIS EXERA II:
Table of Contents

Advertisement

> 10 cm
Figure 4.2
1.
Move the elevator control lever in the opposite direction of the "
direction until it stops.
2.
If necessary, apply a medical-grade, water-soluble lubricant to the insertion
section.
3.
Place the mouthpiece between the patient's teeth or gums, with the outer
flange on the outside of the patient's mouth.
4.
Insert the distal end of the endoscope through the opening of the
mouthpiece, then from the mouth to the pharynx while viewing the
endoscopic image. Do not insert the insertion section into the mouth beyond
the insertion section limit mark.
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
To prevent the patient from breaking one or more teeth, make
sure there are no missing teeth, or teeth that are not
permanently capped, etc., before the examination.
Do not apply olive oil or products containing petroleum-based
lubricants (e.g., Vaseline
products may cause stretching and deterioration of the
bending section's covering.
Do not allow the insertion section to be bent within a distance
of 10 cm or less from the junction of the boot. Insertion
section damage can occur (see Figure 4.2).
Chapter 4 Operation
®
) to the endoscope. These
U"
45