Insertion - Olympus EVIS EXERA II Operation Manual

Duodenovideoscope
Hide thumbs Also See for EVIS EXERA II:
Table of Contents

Advertisement

Chapter 4 Operation
4.1

Insertion

Holding and manipulating the endoscope
Insertion of the endoscope
44
The control section of the endoscope is designed to be held in the left hand. The
air/water and suction valves can be operated using the left index finger. The
UP/DOWN angulation control knob and the elevator control lever can be
operated using the left thumb. The right hand is free to manipulate the insertion
section and the RIGHT/LEFT angulation control knob (see Figure 4.1).
Figure 4.1
Keep the elevator control lever moved all the way in the
opposite direction of the "
withdrawing the endoscope into or from the patient. If the
elevator control lever is moved in the "
operator feels heavy and the forceps elevator is raised while
inserting or withdrawing the endoscope into or from the
patient, this may cause patient injury.
To prevent the patient from biting the insertion section during
an examination, it is strongly recommended that a
mouthpiece be placed in the patient's mouth before inserting
the endoscope.
To prevent the patient from accidentally loosening a dental
prosthesis, make sure that the patient removes it before the
examination.
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
U" direction while inserting or
U" direction until the