Field strength from fixed RF transmitters as determined by an
electromagnetic site survey
compliance level in each frequency range
a)
Field strength from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the
location in which this model is used exceeds the
applicable RF compliance level above, this model
should be observed to verify normal operation. If
abnormal performance is observed, additional
measures may be necessary, such as re-orienting or
relocating this model.
b)
Over the frequency range 150 kHz to 80 MHz, field
strength should be less than 3 V/m.
a)
should be less than the
b)
.
Appendix
87