Treatment screen and clinical setup menus
Parameter
Modes
CPAP
Treatment screen
CPAP
ü
AutoSet
Min
Max
Ramp
ü
Settings
Mode
Set pressure
ü
(CPAP)
Max pressure
Min pressure
Max ramp
ü
Start pressure
ü
EPR
ü
EPR level
ü
Mask
ü
Tube type
ü
Default
AutoSet
ü
ü
4 cm H
O
2
20 cm H
O
2
–
ü
–
ü
10 cm H
O
2
20 cm H
O
ü
2
4 cm H
O
ü
2
30 minutes
ü
4 cm H
O
ü
2
Ramp
ü
1
ü
Nasal
ü
SlimLine
ü
Range
–
–
4–20 cm H
O,
2
0.2 cm H
O
2
increments
Min–20 cm H
O,
2
0.2 cm H
O
2
increments
4–Max cm H
O,
2
0.2 cm H
O
2
increments
Off–45 minutes,
5-minute increments
4–Min pressure
(AutoSet mode),
4–Set pressure
(CPAP mode),
0.2 cm H
O
2
increments
Off / Ramp Only /
Full Time
Off / 1 / 2 / 3 /
Patient
Pillows /
Full Face /
Nasal /
Nasal Ultra
SlimLine /
Standard
Description
Shows the set pressure.
Shows the treatment pressure.
Orange icon shows that device is
ramping up.
Sets the therapy mode depending
on what modes are available on the
device.
Sets the fixed treatment pressure.
Sets the upper limit of treatment
pressure.
Sets the lower limit of treatment
pressure.
Limits the ramp times the patient
may select.
Sets the pressure at the start
of ramp, up to fixed treatment
pressure (CPAP) or minimum
treatment pressure (AutoSet
mode).
Sets when EPR is to occur.
Sets the EPR value.
Selects the type of mask used by
the patient.
Selects the type of air tubing used
by the patient.
19
English