Reassembling The H5I; Maintenance Checklist; Reprocessing The H5I Between Patients - ResMed S9 Elite Clinical Manual

Hide thumbs Also See for S9 Elite:
Table of Contents

Advertisement

Reassembling the H5i

1. Place the plate back onto the base ensuring that the maximum water level mark faces up.
2. Place the tub lid back onto the plate/base ensuring that the centre holes are aligned.
3. Clip all four side latches.
4. Fill the water tub and return the water tub to the H5i.
5. Close the flip lid ensuring that it clicks into place.

Maintenance checklist

ü Inspect the H5i water tub and flip lid seal for wear and deterioration.
ü Replace the water tub if any component is leaking or has become cracked, cloudy or pitted.
ü Replace the flip lid seal if cracked or torn.
ü Clean white powder deposits in the water tub by using a solution of one part household vinegar to 10
parts water.
ü Inspect the maximum water level mark. If cleaning is required, remove by pinching and pushing out
the locks from the plate. Wash in warm water using a mild detergent.

Reprocessing the H5i between patients

The H5i should be reprocessed when used between patients. Cleaning and disinfection instructions are
available from the ResMed website, www.resmed.com under Products and Service & Support. If you
do not have internet access, please contact your ResMed representative.