Replacing the air filter
Replace the air filter every six months (or more often if necessary).
1. Remove the air filter cover from the back of the S9 device.
2. Remove and discard the old filter.
3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
4. Replace the air filter cover.
Notes:
•
Ensure the air filter and air filter cover are fitted at all times.
•
Do not wash the air filter. The air filter is not washable or reusable.
Air filter cover
Air filter
The following filters are available for use with S9 devices:
Filter
Standard (ASMB 160)
Hypo-allergenic (Air Safety Electret100 – electrostatic filter)
Antibacterial filters
Antibacterial filters increase resistance in the air circuit and may affect accuracy of displayed and
delivered pressure, particularly at high flows. ResMed recommends using a filter with a low impedance
(eg, less than 2 cm H
O at 60 L/min).
2
Antibacterial filters are not compatibile with ClimateLine, ClimateLine
electrical connection will not be available.
Efficiency
88% at 7 micron
89.8% at 0.5 micron, bacterial
efficiency of 99.54%.
heated air tubing as the
MAX
27
English