Operating information

The following table describes the operating modes available on the S9 AutoSet and S9 Elite.
Mode
CPAP mode
A fixed pressure is delivered with optional Expiratory Pressure
Relief (EPR). EPR can help ease any discomfort that your
patient experiences when breathing out during treatment.
AutoSet mode
The AutoSet algorithm automatically adjusts pressure in
response to inspiratory flow limitation, snore and apnoea.
AutoSet mode offers optional EPR. You can set the minimum
and maximum allowable treatment pressures.

EPR technology

Designed to make therapy more comfortable, EPR maintains optimal treatment for the patient during
inhalation and reduces the delivered mask pressure during exhalation. You can select EPR in either CPAP
or AutoSet mode.

EPR checklist

ü You can set EPR to Off, Ramp Only, Full Time.
ü You can select an EPR Level of Off, 1, 2, or 3 cm H
ü In CPAP mode you can set the EPR Inhale to Med or Fast.
ü When EPR is enabled, the delivered pressure will not drop below a minimum pressure of 4 cm H
regardless of the settings.
4
S9 Elite
ü
O, or Patient (configurable by the patient).
2
S9 AutoSet
ü
ü
O,
2