WaterMaster
Electromagnetic flowmeter
Abs. Idx /
Variable Name / Brief Description
Rel. Idx
DEVICE_INSTAL_DATE
38 / 22
Device installation date text.
LOCAL_OP_ENA
39 / 23
Blocks editing of values via the local device display, fixed to enabled.
1: Local Operation Enabled
IDENT_NUMBER_SELECTOR
Profile ident currently in use by the device.
Changes to this parameter cause the instrument to reboot.
40 / 24
0: Profile Specific – 0x9700
1: Manufacturer Specific – 0x3431
128: Manufacturer Specific – 0x9740
129: Manufacturer Specific – 0x9741
HW_WRITE_PROTECTION
State of the hardware write protect switch. Enabling the switch disables writing from all
41 / 25
user interfaces.
0: HW Write Protection Disabled
2: HW Write Protection Enabled
FEATURE
42 / 26
Feature information for the device.
COND_STATUS_DIAG
Enables/disables condensed diagnosis.
43 / 27
0: Condensed Diagnosis Disabled
1: Condensed Diagnosis Enabled
PI RESERVED
44 / 28
Parameter reserved for future use.
PI RESERVED
45 / 29
Parameter reserved for future use.
PI RESERVED
46 / 30
Parameter reserved for future use.
PI RESERVED
47 / 31
Parameter reserved for future use.
PI RESERVED
48 / 32
Parameter reserved for future use.
DIAG_WORST_COND
49 / 33
Worst case active diagnostic class.
DIAG_EXT_HISTORY
50 / 34
The device diagnostic history.
UNUSED
51 / 35
Unused parameter.
Table 2.2 Slot 0 – Physical Block (Continued)
IM/WMPBST–EN Rev. B
Access /
Data Type
Size
Store
Usage /
Transport
OctetString
16
S
r,w / C / a
Unsigned 8
1
N
r,w / C / a
Unsigned 8
1
S
r,w / C / a
Unsigned 8
1
D
r / C / a
OctetString
8
N
r / C / a
Unsigned 8
1
S
r,w / C / a
Unsigned 16
2
D
r / C / a
OctetString
6
D
r / C / a
2 Parameter Tables
Default
'
'
Local Operation
Enabled
Manufacturer
Specific – 0x3431
HW Write Protection
Disabled
0
Condensed
Diagnosis Enabled
0x3FFF
0x0,0x0,0x0,0x0,0x
0,0x0
5