Keysight Technologies PXA N9030A Getting Started Manual page 73

Hide thumbs Also See for PXA N9030A:
Table of Contents

Advertisement

Instrument Operating System
Disk Drive Recovery Process
— To re-install the software version you recovered to, follow the process
below.
The following steps load a copy of the instrument software that forces the
program code resident in the instrument's various programmable gate arrays
to be in sync with the restored system software.
Table 3-2
Installing the Software
Step
1 Log out of the default user (instrument):
Start
Log Off
— Select
,
2 At the log in prompt enter:
administrator
— User Name:
Keysight4u
— Password:
3 Navigate to C Drive:
Start
My Computer
— Click
,
Temp
the
folder.
4 Locate the software installer program:
"XSA_Installer_A.XX.XX.exe" or
"MXA_Installer_A.XX.XX.exe".
5 Double-click on the installer.exe.
6 A window appears showing the extraction process.
7 Follow the on-screen prompts to continue the
installation.
When you see the message Programming
WARNING
FPGAs...Do NOT turn off power to the
instrument, be sure to do as it says and DO NOT turn
off the instrument power at this time for ANY reason. If
this process is interrupted the instrument most likely
will need to be sent back to an Keysight Service Center
for servicing before it will be usable again.
Log Off
,
!
C: Drive
,
and open
73
Notes
It may take a minute or two for the
installer to start.
Following the extraction screen,
there may be a short period where
there is no activity on the screen.
The desk top may be displayed for
a short period.