Procedures (Ch1): Ac Mv/Volts (Measure) - GE Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
7. Check the calibration is correct
• Select the TC (measure) function
8. Check the equipment gives a TC temperature that is the
same as the temperature on the reference unit ±0.1°C
(0.2°F)
8.13

Procedures (CH1): AC mV/Volts (measure)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration for the applicable AC function.
• Use the local power supply frequency.
• For the AC mV (measure) function:
• For the AC Volts (measure) function:
06 June 2013
point 1 = 200.0 mVAC,
point 2 = 2000.0 mVAC
point 1 = 2.000 AC,
point 2 = 20.000 VAC
K0541 issue 1
Druck DPI620 Genii
Page 165 of 175