Procedures (CH1/CH2): DC mV/Volts (measure)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a three-
point calibration.
(-FS, Zero and +FS) for the applicable set of ranges:
mV (measure) ranges
200 mV
2000 mV
4. Check the calibration is correct:
• Select the applicable Millivolts or Voltage (measure)
function. (Ref: Section 2.5)
• Apply the input values that are applicable to the
calibration:
mV: -2000, -1000, -200, -100, 0 (short circuit)
mV: 0, 100, 200, 1000, 2000
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Volts (V): 0, 5, 10, 20, 21, 30.
• Check that the error is within limits (Ref: Table 8-4 or
Table 8-5).
06 June 2013
Volts (measure) ranges
20 V
30 V
K0541 issue 1
Druck DPI620 Genii
Page 146 of 175