GE Druck DPI 620 Genii User Manual page 155

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
5. Check the calibration is correct.
• Set-up the equipment as follows:
• Apply the amplitude values that are applicable to the
calibration (Ref: Table 8-10).
• Check the error is within limits (Ref: Table 8-10).
Table 8-10 Amplitude (source) error limits
Amplitud
Volts (V)
06 June 2013
Source frequency = 0 (For DC output);
Waveform = Square
Calibrator
e
uncertainty
0.2
5.0
10.0
20.0
(V)
0.01
0.01
0.01
0.01
K0541 issue 1
Druck DPI620 Genii
Permitted
DPI 620
error (V)
0.1
0.1
0.1
0.1
Page 155 of 175