GE Druck DPI 620 Genii User Manual page 106

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
4. Read the CH2 value and enter this in the meter value text
box using the keypad.
5. Select SET.
6. Repeat steps 3 and 4 with 20 mA selected. This will
calibrate the transmitter's output current.
06 June 2013
Figure 6-9 Enter Calibration Point
K0541 issue 1
Druck DPI620 Genii
Page 106 of 175