User Manual
8.10
Procedures (CH1): Resistance (source)
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration, for each range.
• Range: 0-400 Ω
• Range: 400Ω -2000Ω
• Range: 2kΩ -4kΩ
06 June 2013
K0541 issue 1
Druck DPI620 Genii