Transmitter 9135 - pH/redox measurement
MEASURE
ELECTRODE
TEMP. COMP. :
Select
ELECTRODE
Redox
TYPE :
ISO Pt : 1,5 pH
SLOPE : 95 %
Uiso : 200 mV
Page 40
S/MEASURE Menu
ELECTRODE menu allow to set up the electrode
model.
TEMP.
COMP.
temperature compensation.
ELECTRODE
Choice of the electrode.
• Glass
TYPE
:
• Antimony
• Redox
• Other
ISO Pt
:
XX,XX pH
The isothermal point corresponds to
the pH value which does not vary
according to the temperature.
SLOPE
:
XX %
Indication of the electrode sensitivity
in % of the theoric value (59,16
mV/pH at 25°C).
Uiso
:
XXXX mV
Uiso is the potential of the isothermal
point.
menu
allow
to
set
up
the