Kodak EKTASCAN 160 User Manual page 52

With the kodak x-omat 2000 processor
Table of Contents

Advertisement

Calibration
4-8
Press the Menu Button at the density value to be
13.
changed. The cursor will flash, which will allow you to
change the values. Use the " " or " " to change the
value. Press the Menu Button to accept the change.
Ready
Density Step 1
>0.14 _
Exit
Cancel Change
Menu to change
Use the " " to move the ">" to Exit. Press the Menu
14.
Button.
Repeat steps 12 - 14 for all densities to be changed.
15.
Use the " " to move the ">" to Exit. Press the Menu
16.
Button to return to the Calibration Densities Screen.
Ready
Receive Magazine
Last Calibrate on
13-Dec-1998 07:24
>Exit
Density Values
Compute Calibration
Use the " " to select "Compute Calibration".
17.
Press the Menu Button to complete the calibration.
18.
The calibration information is erased from the screen
display, and is replaced by "Computing Film Model". If
the calibration is not successful, see Chapter 5
"Troubleshooting" for more information.
19.
Use the " " to move the ">" to Exit. Press the Menu
Button 2 times to return to the Main Menu.
Ready
Receive Magazine
Calibration Complete
13-Dec-1998 07:24
>Exit
Density Values
Compute Calibration
5E2234 June 2000