Kodak EKTASCAN 160 User Manual page 51

With the kodak x-omat 2000 processor
Table of Contents

Advertisement

5E2234 June 2000
Use the " " to select "Calibration Data". Press the
10.
Menu Button. The Calibration Data Screen is dis-
played.
Ready
Receive Magazine
Last calibrated on
13-Dec-1998 07:24
Exit Menu
Print Cal Film
> Calibration Data
Use the " " to select "Density Values". Press the
11.
Menu Button. The Calibration Densities Screen is
displayed.
Ready
Receive Magazine
Last Calibrate on
13-Dec-1998 07:24
Exit Menu
>Density Values
Compute Calibration
12.
Use the " " to select each density to be changed.
Press the Menu Button .
Ready
Calibration Densities
>Exit Menu
1) 0.25
2) 0.32
3) 0.69
4) 1.31
5) 1.95
Calibration
4-7