Calibration
Performing the Calibration
4-4
NOTE: This calibration procedure can be per-
formed on a stand-alone Laser Imager or when
docked to a 2000 Processor.
From the Main Display, check the selected Film Des-
1.
tination. If it is not the desired destination, change the
selected Film Destination.
From the Main Display, press the Menu Button. The
2.
Main Menu is displayed.
Ready
>Exit Menu
Access Receive
Access Film Supply
Calibrate Printer
Setup Menu
Film Destination
Use the " " to move the ">" to "Calibrate Printer".
3.
Press the Menu Button. The Calibrate Printer Screen
is displayed.
Ready
Exit Menu
Access Receive
Access Supply
>Calibrate Printer
Setup Menu
Film Destination
The Calibrate Printer Screen displays the calibration
time and date for the currently selected destination.
5E2234 June 2000