Kodak EKTASCAN 160 User Manual page 46

With the kodak x-omat 2000 processor
Table of Contents

Advertisement

Calibration
Introduction
Calibration Screens
4-2
Calibration is the method of ensuring consistent image
appearance by compensating for variations in laser
power, differences between film emulsions and other
process variables.
IMPORTANT: When the Laser Imager is docked
to a 2000 Processor, separate calibration is
required for the docked 2000 Processor and the
Receive Magazine.
You will do the following during calibration:
1.
If the 2000 Processor is docked to the Laser Imager,
select "Film Destination".
Print and process the calibration film.
2.
Read the densities of the film and enter the density
3.
values.
4.
Compute the Calibration.
Screen
Calibrate Printer
Calibration Data
Calibration Densities Allows viewing and modification
Function
Prints the calibration film and
allows access to calibration data
Allows viewing and modification
of Laser Imager calibration data
of Laser Imager calibration den-
sities
5E2234 June 2000