Sterilization Procedures; Handpiece; Handpiece Cradle; Accessories - Welch Allyn Kreativ Mach 5.0 Instructions For Use Manual

Table of Contents

Advertisement

6.

Sterilization Procedures

WARNING: All Kreativ Mach 5.0 and Mach 5.0 PLUS accessories which
come into contact with the patient should be sterilized prior to each use. The
handpiece cradle should be wiped with disinfectant prior to each patient use.

Handpiece

Sterilization of the cutting handpiece and nozzles can be accomplished by using any
conventional dental autoclave. The handpiece and nozzles are fully autoclaveable in a
dry heat, steam, or chemical autoclave. This is recommended prior to each use to
ensure that the handpiece and nozzles do not become a source for cross contamination
of disease or infection.
Prior to sterilizing, the SuperSonic Turbo
cleared of all residual powder by blowing clean, dry air through the handpiece.

Handpiece Cradle

Always wipe down the handpiece cradle in the remote control unit before each patient
use to avoid contamination of the handpiece from an earlier patient session.
WARNING! Do not attempt to disassemble the SST Handpiece Assembly.

Accessories

The handpiece/nozzle hose assembly of the Kreativ Mach 5.0 and Mach 5.0 PLUS
should be sheathed or wiped down in accordance with standard dental procedures.
(SST) Handpiece Assembly should be
Page 24
Chapter
6
990008 Rev. H