Maximum and Minimum Pulse Rate Ranges
The maximum and minimum pulse rate ranges are as follows:
Performance Specifications
Patient Population
The Welch Allyn Spot Vital Signs is designed for use with adult and pediatric patients. Welch
Allyn defines a pediatric patient as 29 days old or more.
THE WELCH ALLYN SPOT VITAL SIGNS IS NOT INTENDED FOR USE ON NEONATES .
Welch Allyn defines neonates as children 28 days or less of age, born at term (37 weeks gesta-
tion or more), otherwise up to 44 gestational weeks.
Cuff Pressure Range
0 mmHg - 300mmHg
Initial Cuff Inflation
160 mmHg
Systolic Determination
Maximum:
250 mmHg
Minimum:
60 mmHg
Diastolic Determination
Maximum:
160 mmHg
Minimum:
30 mmHg
Blood Pressure Accuracy
Blood pressure accuracy meets or exceeds SP10-1992 AAMI standards for non-invasive blood
pressure accuracy (AAMI standard: ± 5 mmHg mean error, 8 mmHg standard deviation). Blood
pressure accuracy is validated for pressure measurement using the upper arm only.
Blood Pressure Determination Time
Typical:
20 to 45 seconds
Maximum:
165 seconds
Pulse Rate Determination (using SpO
Maximum:
245bpm
Minimum:
25bpm
16
Measurement
Using BP determination
Using SpO
determination
2
Maximum
Minimum
200 bpm
40 bpm
245 bpm
25 bpm
determination)
2
Welch Allyn