Setting The Low Voltage Threshold; Selecting A Test Mode (Jars Only); Selecting The Number Of Posts Per Jar - Midtronics Celltron Ultra Instruction Manual

Universal stationary battery analyzer
Hide thumbs Also See for Celltron Ultra:
Table of Contents

Advertisement

Celltron ULTRA
Setting the
LOW VOLTAGE
1. Press the ARROW (or use the numeric keypad) to highlight LOW VOLTAGE, and press
ENTER .
2. Press the ARROW to scroll to the number of volts.
3. Press ENTER to accept.
Range
1.000 to 2.750 V for a 2 V jar
2.000 to 5.500 V for a 4 V jar
3.000 to 8.250 V for a 6 V jar
4.000 to 11.000 V for an 8 V jar
Selecting a test mode
The default is JARS ONLY.
1. Press the ARROW keys to highlight JARS ONLY, JARS & 1 STRAP, or JARS & 2 STRAPS in
the STRING INFO list, and press ENTER .
2. Press the ARROW keys (or use the numeric keypad) to check the box next to JARS ONLY,
JARS & 1 STRAP, or JARS & 2 STRAPS.
3. Press ENTER to accept.
Selecting the number of
The default is 2.
1. Press the ARROW keys to highlight POSTS PER JAR, and press ENTER .
2. Press the ARROW keys (or use the numeric keypad) to scroll to the number of posts.
3. Press ENTER to accept.
threshold
Table 6: Low voltage threshold ranges and defaults
Default
2.1 V
4.2 V
6.3 V
8.4 V
(JARS ONLY)
POSTS PER JAR
Chapter 5: Selecting test parameters
Range
5.000 to 13.800 V for a 10 V jar
6.000 to 16.500 V for a 12 V jar
7.000 to 19.250 V for a 14 V jar
8.000 to 22.000 V for a 16 V jar
31
Default
10.5 V
12.6 V
14.7 V
16.8 V