Setting and Default are sensor specific, depending on the setting in Process value and
depending on the Units setting.
Setting
Default
Related
Path
3.4.3.3.7 Damping value
Description
Time constant of damping filter for frequency output signal.
Setting
Default
Related
Path
3.4.3.3.8 Fail-safe behaviour
Description
Frequency output reaction in case of a fault.
Setting
Default
Related
Path
3.4.3.3.9 Fail-safe minimum time
Description
Minimum time the output stays in fail-safe mode.
Setting
Default
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
-108000000.00 kg/h to +10800000000.00 kg/h
0.00 kg/h
Setup → Inputs and outputs → Channel 4 - input/output → Frequency output → Lower range
value
0.0 s to 100.0 s
0.0 s
Setup → Inputs and outputs → Channel 4 - input/output → Frequency output → Damping
value
Minimum frequency; Maximum frequency; Last valid value; Current valid; fail-safe value
Last valid value
Setup → Inputs and outputs → Channel 4 - input/output → Frequency output → Fail-safe
behaviour
0 s to 100 s
0 s
Setup
3.4 Inputs and outputs