Siemens FST020 Function Manual page 32

Sitrans f series ultrasonic flowmeters ip65 nema 4x
Table of Contents

Advertisement

Setup
3.1 Sensor settings
Related
Path
3.1.7.2.3 Sensor spacing
Description
Sensor spacing - Distance between the front faces of the sensors for path 1.
Setting
Default
Related
Path
3.1.7.2.4 Zero point adjustment
Description
Zero point adjustment. Can only be changed via the Sensor Wizard.
Note
Important
Please stop all flow in the pipe and perform this function. The flow velocity is measured prior
and after the function has been applied.
Setting
Default
Related
Path
3.1.7.2.5 Serial number
Description
Serial number of sensor (clamp-on)
Setting
Default
30
Setup → Sensor settings → Path settings → Path 1 → Path geometry
-100.0 m to 100.0 m
0.0 m
Setup → Sensor settings → Path settings → Path 1 → Sensor spacing
5 to 300
10
Setup → Sensor settings → Path settings → Path 1 → Zero point adjustment → Time duration
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA