Maintenance
Replace Fuses
Extra primary and secondary fuses are included in the accessory box.
NOTE! Replace only with fuses of specified type and rating.
Primary Fuses
1.
Disconnect the mains
power cord.
2.
Open the fuse/voltage se-
lector cover using a screwdriver.
Insert the screwdriver at the
top center and apply pressure.
3.
Take out the fuse holders
from the instrument and re-
move the old fuses.
4.
Place new fuses in the fuse
holders and install them in both
the left- and right-hand posi-
tion in the instrument.
5.
Close the cover.
6.
Connect the mains power
cord.
P a g e 1 1 . 2 0
NOTE! Phadia 100 must only be connected to a
grounded power outlet (protective ground).
Phadia 100 - User Manual
12-3501-20/12