Installing/Mounting
4.1
Application planning
Application planning of energy calculator consists of three steps:
1. Ambient conditions check.
2. Check of application settings.
3. Device installation, see Installation of device (Page 17).
4.2
Ambient conditions
Specifications
SITRANS FUE950 energy calculator suitable for indoor and outdoor installations.
● Temperature specifications:
● Enclosure rating:
● Environment classes:
SITRANS FUE950
Operating Instructions, 01/2010, SFIDK.PS.022.F4.02
– Ambient temperature: 0 ... +55 ºC (+32 ... +131 ºF)
– Storage: –25 °C ... +70 °C (-13 ... +158 °F)
– IP54
– EN1434 class E1 / M1 (electromagnetic / mechanical)
Note
Ensure calculator is installed sufficiently far away from possible sources of
electromagnetic interference (switches, electric motors, fluorescent lamps, etc.).
Ensure that temperature and flow specifications indicated on device type plate / label
will not be exceeded.
4
15