Installing/Mounting; Application Guidelines; Mounting The Transmitter - Siemens SITRANS FUS1010 Quick Start Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS FUS1010:
Table of Contents

Advertisement

Installing/Mounting

2.1

Application Guidelines

Basic Requirements
● Determine pipe material and dimensions.
● Avoid vertical pipes flowing in a downward direction.
● Avoid installation of sensors on the top and bottom of horizontal pipes, if possible.
● Select a location with the longest straight run of pipe.
● Identify upstream piping configuration (elbow, reducer, etc.).
● Pipe surface should be smooth and, if necessary, free of paint.
● Avoid pressure reduction components upstream.
● Avoid mounting on or near weld seams.
● Pipe must be full to achieve proper operation.
2.2

Mounting the Transmitter

Hazardous Voltage
May cause death or serious personal injury.
Disconnect power before working on this product.
Wall Mounting
The transmitter can be mounted on any wall surface including wood, metal or concrete. Use
the appropriate bolts and screws as needed for your mounting application and adhere to
local codes. (See figure below for mounting bracket locations.)
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
WARNING
2
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fuh1010Sitrans fue1010

Table of Contents