PNA-L Series test set and power configuration options

The PNA-L is an integrated vector network analyzer featuring a
built-in S-parameter test set, one synthesized source used for
device stimulus, a solid state drive, USB interfaces, and a 10.4"
LCD touch screen display. The N5239A, N5231A and the N5232A
have 50 ohm, ruggedized 3.5 mm (m) test ports. The N5234A
and N5235A have 50 ohm, ruggedized 2.4 mm (m) test ports.
Included with each instrument is a mouse, keyboard (U.S. style), .
For one day of on-site productivity assistance (not included with
instrument purchase), request quantity 1 each PS-S20-PNA.
A test set and power configuration option is mandatory, choose
one of the following:
2 ports, single source, base configuration (Option
200)
The 2-port with base configuration has no front-panel access
loops.
R1
A
Test port 1
2 ports, single source, with configurable test set
and source attenuators (Option 216)
The 2-port test set comes with a configurable test set and
source attenuator at each port. The configurable test set adds six
front-panel access loops. The loops provide access to the signal
path between (a) the source output and the reference receiver,
(b) the source output and directional coupler thru arm, and (c) the
coupled arm of the directional coupler and the port receiver at all
ports. The source attenuators are 60 dB in 10 dB steps.
60 dB
R1
A
Test port 1
Find us at www.keysight.com
LO
Source
To receivers
R2
B
Test port 2
LO
Source
To receivers
60 dB
R2
Test port 2
4 ports, single source, base configuration (Option
400)
The 4-port base configuration has no front-panel loops, and is
available only on the N5231A and N5232A.
R
4 ports, single source, with configurable test set
and source attenuator (Option 416)
The 4-port test set comes with a configurable test set and one
source attenuator to be shared with all ports. The configurable
test set adds nine front-panel access loops. The loops provide
access to the signal path between (a) the source output and the
reference receiver, (b) the source output and directional coupler
thru arm at all ports, and (c) the coupled arm of the directional
coupler and the port receiver at all ports. The source attenuator
is 60 dB in 10 dB steps. This configuration is available on only
N5231A and N5232A.
B
Source
A
B
Test port 1
Test port 2
Test port 3
Source
60 dB
R
A
B
Test port 1
Test port 2
Test port 3
LO
To receivers
C
D
Test port 4
LO
To receivers
C
D
Test port 4