Memory Operations
Datalog

Running Procedure Files (DPI 615 instruments only)

To run a procedure, make sure that the instrument is set to Store mode, Data Log (see
page 36), and proceed as follows:
After selecting F1, proceed by entering the User ID and Serial Number and then select F1
(Continue) and follow the on-screen procedural instructions:
When the test procedure for a given UUT has been completed, the result of running the
first test is stored as an AS FOUND file. This file cannot be overwritten. Any subsequent
tests on the device are stored as an AS LEFT file that is overwritten each time the
procedure on this device is run.
When recalling the results of a procedure, the choice of AS FOUND or AS LEFT is provided
(refer to page 35).
K0430 Issue No. 2