Ambient Temperature Channel - GE Druck DPI 610 User Manual

Intrinsically safe portable pressure calibrator
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Calibration

Ambient Temperature Channel

Use the following procedure for calibrating the ambient temperature measurement
channel.
(1)
Switch the instrument on, enter the CALIBRATION mode and select
TEMPERATURE:
(2)
Allow the instrument's temperature to stabilize in the calibration environment for
at least one hour.
(3)
Read the environmental temperature on a calibrated digital thermometer and
enter the recorded value as shown below. Example shown for a measured
ambient temperature of 21.5° Celsius.
Note: Only one temperature point is required.
(4)
Press the ENTER key to accept the calibration. Press the EXIT key once to return
to the calibration menu or four times to quit the CALIBRATION and SETUP modes.
K0430 Issue No. 2
54