Environment Conditioning Requirements

Environment conditioning considerations include temperature, humidity,
altitude, atmosphere, airborne dust and exhaust venting.
Temperature and Humidity
The 7700 ICP- MS is specified for operation under the conditions outlined in
Table
Table 9
Parameter
Temperature
Humidity
Altitude
Atmosphere
Table 10
Table 10 Heat Output
Item
1
2
3
4
5
Item
6
7
8
A maximum of 2,500W is removed from the 7700 ICP-MS mainframe via the cooling water
N O T E
and extraction duct. Approximate values. Refer to the specific product information for
details.
The ambient temperature around the heat exchanger must not exceed 30°C for normal
ICP-MS operation.
G3280-90010 (Rev.B) Agilent 7700 Series ICP-MS Site Preparation Guide
9.
Temperature and Humidity Specifications
Specification
15-30°C (59-86°F) < 2°C/h change and total change should be < 5°C
20-80%
Up to 2,000m
Non Condensing; Non Corrosive
lists the heat output of the 7700 ICP- MS and other components.
Product
7700 ICP-MS
Foreline Pump
PC/Monitor
Heat Exchanger (50/60Hz)
Agilent Chiller (G3292A)
Product
Heat Exchanger (50/60Hz)
Agilent Chiller (G3292A)
Extraction Vent
Heat Dissipated (W)
2,900
500
430
2,000 (Maximum)
50/60Hz
3,200 (Maximum)
Heat Absorbed (W)
1,300 (Maximum)
50/60Hz
1,300 (Maximum)
1,200 (Maximum)
23