Maico ERO-SCAN Operating Instructions Manual page 60

Hide thumbs Also See for ERO-SCAN:
Table of Contents

Advertisement

Operating Instruction
ERO•SCAN
Screening and Diagnostic Version
Message
Limit
Error
Memory almost full
Memory Full!
Power Low!
Printer Error
Printer Paper Out!
Time/Date Error
operating_instructions_8106556-6_EROSCAN_e_15a
®
Solutions
 Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test.
 Cycle instrument power.
 Contact MAICO for service if the problem persists.
 Saved tests are within 5 tests of the maximum limit. Print or
transfer test result to avoid interruption in testing.
 The maximum saved test limit is reached. The user will need to
clear the memory before any additional tests can be performed.
 The battery charge level is too low for operation. The user must
charge the battery before additional tests can be performed.
 Indicates a problem with the printer. Check the printer status.
 Reset the printer or cycle the printer power.
 Replace the paper roll.
 The clock is checked during power on to ensure it has not lost
time and been reset. In the case of clock reset, this message is
shown. The user should set the correct date/time.
53
8106556-5 02/15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents