19.4 Setting the Number of Frequencies for PASS

Display 36

19.5 Reset Protocol

Display 37

19.6 Save Protocol

Display 38
operating_instructions_8106556-6_EROSCAN_e_15a
®
The number of frequencies for determining a PASS can be set
from 0 to 6. If the setting is on 0, then no indication of
PASS/REFER will be made. This setting is used in conjunction
with the PASS SNR to set the criteria for the overall test
PASS/REFER indication. For example, if the PASS SNR is set to 4
dB and the number of frequencies for PASS is set to 3 then the
test must contain at least 3 frequencies where the emission is at
least 4 dB above the noise to indicate a PASS.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 36).
NOTE: Diagnostic protocols display the SNR graph with purple
bars. Once the "Number of Frequencies for PASS" is set above
zero, green bars display to identify this setting.
Selecting the RESET arrow key in the Reset Protocol menu
(Display 37) will return the selected protocol settings to their
original factory settings. Press the NEXT key to exit.
NOTE: This does not affect the instrument settings or the
settings of any other protocol.
Once all of the settings have been selected for the protocol,
these settings can be saved by selecting the SAVE keys
(Display 38). Press the DONE key to exit.
Selection of DONE before saving will not save the selections
made to the protocol.