Operating Instruction
ERO•SCAN
Screening and Diagnostic Version
Message
Attach
Probe
BT Device
Not Found
BT Error
#xxx
BT Not
Configured
Device Not
Responding
Due for Service
Fit Error
Cannot Obtain L
Fit Error
Too High
Fit Error
Too Low
operating_instructions_8106556-6_EROSCAN_e_15a
®
Solutions
Probe not detected. Check that the probe connector is fully
seated in the socket.
Disconnect and reconnect the probe.
Cycle instrument power.
Contact MAICO for service if the problem persists.
Paired to Printer:
o Check that the printer is turned on.
o Move closer to the printer.
o Try again.
Paired to PC Computer or Bluetooth
o Check that the serial port is open. Establishing the serial
port is handled by the PC and/or the software, not by
the ERO•SCAN
Check BT device (printer or PC) status.
Attempt to connect to BT device again.
Contact MAICO if the problem persists.
Printing has been attempted, but no BT device is paired with
®
the ERO•SCAN
The printer is not responding to queries from the instrument.
Check printer status.
Awaken printer from sleep mode.
Charge printer battery if necessary.
Indicates calibration of instrument is recommended. Message
will appear upon the calibration due date set in the device.
Message appears during device startup once per day.
For a DP test, the desired level (L1 or L2) cannot be obtained
within allowable limits. User should refit the probe and retry
the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
For a DP test, the level of the calibration tone is too high. User
should refit the probe and retry the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
For a DP test, the level of the calibration tone is too low. User
should refit the probe and retry the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
®
®
instrument.
Instrument. Establish Bluetooth
52
dongle:
®
pairing.
8106556-5 02/15