Installation and connection

Information on installation

Caution!
• The analyzer may only be installed by trained experts
in accordance with this instruction manual
and as per applicable local and national codes.
• Be sure to observe the technical specifications and
input ratings.
• Be sure not to notch the conductor when stripping the
insulation.
• Before connecting the analyzer to the power supply,
make sure that its voltage lies within the range
20.5 to 253 V AC/DC.
• When commissioning, a complete configuration must be
carried out by the system administrator.
The terminals are suitable for single wires and flexible leads up to
2.5 mm
2
(AWG 14).

Terminal assignments

Fig. 6: Terminal assignments APT4000PH
12
1 ESD shield covering the signal inputs (Screw off for assembly)
2 Terminals for temperature probe and outer shield
3 Terminals for electrode
4 Connection of power supply
Fig. 7: Information on installation, rear side of analyzer
Division 2 wiring
APT4000PH
4
Note: The cable shield must end under the ESD shield.
(Cut lines if required)
The connections to the analyzer are incendive and
must be installed in accordance with the National
Electric Code (ANSI-NFPA 70) Division 2 hazardous
(classified) location incendive wiring techniques.
3
1
2
13