Output 1 - Honeywell APT4000PH User Manual

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration

Output 1

Select type of electrode. Measurement procedure.
Menu group
code
Display
Output 1
o1.
32
Select menu item
Select measured variable
Select electrode type
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:
Press CONF, then ENTER
APT4000PH
Code Display
Action
o1.
Select configuration
(Press CONF.)
Enter mode code "1200"
(Select position with
edit number with
When the display reads "1200",
After correct input a wel-
press ENTER to confirm.)
come text (ConF) is dis-
played for approx. 3 s
The APT4000PH is in Hold mode
(Hold icon is on).
Select measured variable pH/ORP
Select with
Proceed with ENTER
Only with pH selected:
Select electrode type:
• ISFET electrode
• Glass electrode
Select with
Proceed with ENTER
Note: Characters represented in gray are flashing and can be edited.
Choices
key and
key.
pH/ORP
key
FEt
(Glas)
key
33